首页 / 资料库 / 文献详情

Radon Measurement Using a Liquid Scintillation Spectrometer

Taeko KogaHiroshige MorishimaHiroshi KawaiSohei KondoMasaki MifuneMasanobu KonishiChiaki Shirai

1992Institutional Repositories DataBase (IRDB)Physics and Astronomy被引 8开放获取

出版方页面 →

摘要

A convenient radon detecting device for the purpose of estimating natural radiation exposure is described. The a radioactivity of radon gas adsorbed in fine active carbon particles exposed to air is measured with a liquid scintillation spectrometer (Packard-PICO-RAD system). Its detection limits are 2 mBq/ E in air and 0.5 Bq/ E in water with an accuracy of about 10%. Radon concentrations at Misasa hot springs in Tottori prefecture were measured using this method. They were 0.16--7.7 Bq/£ in a bath room and 0.057--0.36 Bq/£ outdoors. Radon concentrations of the hot springs were 82--1, 700 BqjE.

引用本文(GB/T 7714)

Taeko Koga, Hiroshige Morishima, Hiroshi Kawai, 等. Radon Measurement Using a Liquid Scintillation Spectrometer[J]. Institutional Repositories DataBase (IRDB), 1992.