首页 / 资料库 / 文献详情

Time Performance Study of Plastic Scintillator Detector with SiPM Readouts

Bingqian ZhouXueheng ZhangFang FangD. YanShu-Weng TangYuhong YuShitao WangYongjie Zhang

2020原子核物理评论Physics and Astronomy被引 1

出版方页面 →

摘要

In this paper the time performance of a plastic scintillator detector with an area of 10 cm×10 cm and read out by multiple silicon photomultipliers (SiPMs) at both ends was studied. The results tested with a 239Pu source are shown as: (1) the time resolutions became better gradually with the increase of the quantities of SiPMs connected in series; (2) when the number of 12 SiPMs was fixed, the time resolutions got worse as the number of parallel branches were increased; (3) the detector time resolutions can be effectively improved by using a fast-time plastic scintillator and adding its thickness; (4) to improve the positional uniformity of the time resolution, the scintillator with larger dimensions than beam spots should be used; (5) a resolution less than 131 ps was achieved with a 0.1 cm thickness EJ232 scintillator detector read out on each side by 12 SiPMs conncted in series. This study has importance significance for the upgrading of the start detector at RIBLL2.

引用本文(GB/T 7714)

Bingqian Zhou, Xueheng Zhang, Fang Fang, 等. Time Performance Study of Plastic Scintillator Detector with SiPM Readouts[J]. 原子核物理评论, 2020.

引文网络

参考文献与被引分析加载中…

DOI:https://doi.org/10.11804/nuclphysrev.37.2019cnpc53

本站仅收录题录与摘要供学习参考,全文版权归属出版方;如有侵权请联系我们删除。