非分散红外气体分析系统中检测灵敏度的提高方法研究

Youwen SunWenqing LiuShimei WangShuhua HuangXiaoman Yu

2011Chinese Optics LettersEnvironmental Science被引 4

出版方页面 →

摘要

A method of interference correction for improving the sensitivity of non-dispersive infrared (NDIR) gas analysis system is demonstrated. Based on the proposed method, the interference due to water vapor and carbon dioxide in the NDIR NO analyzer is corrected. After interference correction, the absorbance signal at the NO filter channel is only controlled by the absorption of NO, and the sensitivity of the analyzer is improved greatly. In the field experiment for pollution source emission monitoring, the concentration trend of NO monitored by NDIR analyzer is in good agreement with the differential optical absorption spectroscopy NO analyzer. Small variations of NO concentration can also be resolved, and the measuring correlation coefficient of the two analyzers is 94.28%.

引用本文(GB/T 7714)

Youwen Sun, Wenqing Liu, Shimei Wang, 等. 非分散红外气体分析系统中检测灵敏度的提高方法研究[J]. Chinese Optics Letters, 2011.

引文网络

参考文献与被引分析加载中…

DOI:https://doi.org/10.3788/col201109.060101