首页 / 资料库 / 文献详情

[Bamboo carbon as sorbent for the separation and preconcentration of bismuth with detection by HG-AFS and ICP-MS].

Mingli ChenLiming ShenJianhua Wang

2011PubMedChemistry被引 2

出版方页面 →

摘要

A procedure for the separation and preconcentration of bismuth was developed in a sequential injection system by employing bamboo carbon as sorbent. The detection was facilitated by both hydride generation atomic fluorescence spectrometry and inductively coupled plasma mass spectrometry. With a sample volume of 1 mL, a detection limit of 13 ng x L(-1) was obtained, along with a precision of 0.9% (0.3 microg x L(-1), n = 9) with detection by HG-AFS, and a detection limit of 10 ng x L(-1) along with a precision of 2.8% (0.3 microg x L(-1), n = 5) was achieved with detection by ICP-MS. The present system was validated by analyzing a certified reference material of river sediment (CRM 320), and spiking recovery of bismuth in human whole blood was performed with hydride generation atomic fluorescence spectrometry. No significant difference was identified in the results of bismuth detection in blood samples by hyphenating the present solid phase extraction system with detection by hydride generation atomic fluorescence spectrometry and inductively coupled plasma mass spectrometry.

引用本文(GB/T 7714)

Mingli Chen, Liming Shen, Jianhua Wang. [Bamboo carbon as sorbent for the separation and preconcentration of bismuth with detection by HG-AFS and ICP-MS].[J]. PubMed, 2011.

引文网络

参考文献与被引分析加载中…

本站仅收录题录与摘要供学习参考,全文版权归属出版方;如有侵权请联系我们删除。