首页 / 资料库 / 文献详情

角膜压平面积光学测量探头的开发

Xueyong ZhangJianguo MaRongsheng LuYibao LiZhen TangGuohong Liu

2010Chinese Optics LettersMedicine被引 2开放获取

出版方页面 →

摘要

The optical probe in tonometers is a key component in measuring the flattened diameter or area of the ocular cornea. A new kind of optical probe for the direct measurement of the flattened area of the ocular cornea is presented. The optical probe uses the cone prism with a modulating flake of light intensity as its measuring body. The test results on simulated eyeballs with different radii of curvature of the ocular cornea show that there is a linear relation between the flattened area of the ocular cornea and the normalized current. The optical probe, which is more compact compared with existing optical probes and easily configured with its coaxiality of optical parts, may be an excellent probe for constructing a low-cost, miniaturized applanation tonometer.

引用本文(GB/T 7714)

Xueyong Zhang, Jianguo Ma, Rongsheng Lu, 等. 角膜压平面积光学测量探头的开发[J]. Chinese Optics Letters, 2010.

引文网络

参考文献与被引分析加载中…

DOI:https://doi.org/10.3788/col20100801.0103

本站仅收录题录与摘要供学习参考,全文版权归属出版方;如有侵权请联系我们删除。