首页 / 资料库 / 文献详情

微型DPSA-1仪-SPCE微分电位溶出法同步快速检测茶叶中的铅、镉、铜

赵广英沈颐涵

2010Engineering被引 1

出版方页面 →

摘要

Mini DPSA-1 device, coupled to screen-printed carbon electrodes (SPCE), was used for simultaneous rapid detection of lead, cadmium and copper in tea by differential potentiometric stripping analysis (DPSA). The optimized experimental parameters were as follows, Hg(superscript 2+) concentration-3×10^(-4) mol/L, supporting electrolyte-1.5 mol/L pH4.0 NH4Cl, deposition potential-1.1 V, final potential-0.2 V, deposition time-200 s. The potentiometric stripping response for lead, cadmium and copper following 200 s deposition was linear over the concentration range examined (20~840μg/L) with r values of 0.99749, 0.99740 and 0.99612 and detection limit of 1.1 2μg/L, 1.09μg/L and 1.23μg/L (S/N=3), respectively. The interferences of various metal ions were also examined in the determination of lead, cadmium and copper. No significant difference existed after the rapid detection results were compared with GB method using t test. In comparing with conventional detection technique, this detection method for lead, cadmium and copper in tea, using mini DPSA-l instrument, coupled to SPCE by DPSA, was a cheaper rapid detection technique. It could achieve a high percentage of sample screening and be applied to the simultaneous rapid detection of lead, cadmium and copper in other food and environmental samples.

引用本文(GB/T 7714)

赵广英, 沈颐涵. 微型DPSA-1仪-SPCE微分电位溶出法同步快速检测茶叶中的铅、镉、铜[J]. 未知来源, 2010.

引文网络

参考文献与被引分析加载中…

DOI:https://doi.org/10.13305/j.cnki.jts.2010.01.010

本站仅收录题录与摘要供学习参考,全文版权归属出版方;如有侵权请联系我们删除。